The Effect of Tissue Physiological Variability on Transurethral Ultrasound Therapy of the Prostate

SUOMI, V.; TREEBY, B.; JAROŠ, J.; SAUNAVAARA, J.; KIVINIEMI, A.; BLANCO, R. The Effect of Tissue Physiological Variability on Transurethral Ultrasound Therapy of the Prostate. In 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). Honolulu, HI: Institute of Electrical and Electronics Engineers, 2018. p. 5701-5704. ISBN: 978-1-5386-3646-6.
Typ
článek ve sborníku konference
Jazyk
anglicky
Autoři
Suomi Visa, FIT (FIT)
Treeby Bradley
Jaroš Jiří, prof. Ing., Ph.D., UPSY (FIT)
SAUNAVAARA, J.
KIVINIEMI, A.
BLANCO, R.
Abstrakt

Therapeutic ultrasound is an established modality which can be used for minimally invasive treatment of prostate cancer. Computational simulation models were used to study the effect of natural physiological variations in tissue parameters on the efficacy of therapeutic ultrasound treatment in the prostate. The simulations were conducted on a clinical ultrasound therapy system using patient computed tomography (CT) data.
The values of attenuation, perfusion, specific heat capacity and thermal conductivity were changed within their biological ranges to determine their effect on peak temperature and thermal dose volume. Increased attenuation was found to have the biggest effect on peak temperature with a 6.9% rise. The smallest effect was seen with perfusion with 0.2% variation in peak temperature. Thermal dose was mostly affected by specific heat capacity which showed a 20.7% increase in volume
with reduced heat capacity. Thermal conductivity had the smallest effect on thermal dose with up to 2.1% increase in the volume with reduced thermal conductivity. These results can be used to estimate the expected interpatient variation during the therapeutic ultrasound treatment of the prostate.

Klíčová slova

Ultrasonic imaging, Thermal conductivity, Heating systems, Conductivity, Computational modeling, Attenuation, Probes

URL
Rok
2018
Strany
5701–5704
Sborník
40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Konference
40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
ISBN
978-1-5386-3646-6
Vydavatel
Institute of Electrical and Electronics Engineers
Místo
Honolulu, HI
DOI
EID Scopus
BibTeX
@inproceedings{BUT155005,
  author="SUOMI, V. and TREEBY, B. and JAROŠ, J. and SAUNAVAARA, J. and KIVINIEMI, A. and BLANCO, R.",
  title="The Effect of Tissue Physiological Variability on Transurethral Ultrasound Therapy of the Prostate",
  booktitle="40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)",
  year="2018",
  pages="5701--5704",
  publisher="Institute of Electrical and Electronics Engineers",
  address="Honolulu, HI",
  doi="10.1109/EMBC.2018.8513548",
  isbn="978-1-5386-3646-6",
  url="https://ieeexplore.ieee.org/document/8513548"
}
Soubory
Projekty
IT4Innovations excellence in science, MŠMT, Národní program udržitelnosti II, LQ1602, zahájení: 2016-01-01, ukončení: 2020-12-31, ukončen
Výzkumné skupiny
Pracoviště
Nahoru